MODULE 3: DAILY START UP AND CHECKS
3-30 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Check High-Concentration Waste Container (If used)
WARNING: Potential Biohazard. Refer to Appendix, Hazards
of this manual.
CAUTION: Chemical Hazard. Refer to Appendix,
Hazards of this manual. Also refer to product-specific
Material Safety Data Sheets (MSDS).
CAUTION: The high-concentration waste container is
heavy when full. Use care when handling the container
to reduce the risk of injury.
1. Ensure that the System status is READY.
2. Disconnect the float switch cable from the
High-Concentration Waste Container cap by unscrewing the
locking ring and disconnecting the cable.
3. Place an absorbent towel next to the container to catch any drips
from the container cap.
4. Unscrew the High-Concentration Waste Container from the cap,
being careful not to kink the tubing.
5. Place the cap (with the attached tubing) on the absorbent towel.
6. Empty the Waste Container according to local waste handling
procedures.
7. Screw the Waste Container onto the cap, being careful not to kink
the tubing.
8. Reconnect the float switch cable by inserting the cable and
tightening the locking ring.
!
!
• If local regulations permit, High-
concentration Waste may be drained
to a floor drain.
• Instruments in the training do not
have a waste container; therefore,
this will not be demonstrated.
Cable
Locking Ring