Daily Checks
AEROSET
®
Customer Training Guide 3-31
94858-103 — July 2003
Check Refrigerator Condensation Pan (If used)
1. Locate the Refrigerator Condensation Pan underneath the rear of the
Analyzer just to the right of the water supply/drainage panel.
Figure 3.8: Refrigerator Condensation Pan
2. Pull the pan out, keeping it level to prevent spills.
3. Empty the water collected in the condensation pan.
NOTE: This water is condensation from the reagent
refrigeration units and is not considered potentially
biohazardous.
4. To replace, align the center of the pan with the center of the large
vent on the rear of the Analyzer.
5. Slide the pan underneath the Analyzer just far enough so that it does
not extend beyond the Analyzer.
• May be drained to a floor drain if
diameter of drain is ≥ 3.5” and
height is ≤ 3”.
Refrigerator
Condensation
Pan