MODULE 4: RUN OPTIONS
4-12 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Additional Options on the Order Samples Screen
Figure 4.5: Order Samples Screen
Sample Replicates
Enter the number of replicates desired in the Replicates field. Up to 5
replicates can be entered.
Example: If 3 replicates are entered, the System will dispense and
analyze each selected assay 3 times.
Manual Dilution
Enter a multiplier in the Manual Dil field, if an off-line dilution was
done. The multiplier will be used to correct the final result for all assays
ordered, both photometric and potentiometric.
Example: If a 1:5 dilution was performed, enter a 5 in this field. The
System will multiply the final result by 5 to correct for the dilution.
• Manual dilution field holds 3 digits.