Sample Order Options
AEROSET
®
Customer Training Guide 4-13
94858-103 — July 2003
Onboard Dilution
The measurement type for each assay can be selected. The options are:
NOTE: Configuration of these options is discussed in Module 8,
Configuration.
To define the measurement type for each assay, first select the type and
then select the assay button. The measurement type is visually displayed
as a colored line on the assay button.
NOTE: The AEROSET System does not perform onboard
dilutions of samples for ICT™ assays. Therefore, if an option is
selected, the underline remains green
.
CAUTION: The Operator must verify the correct
dilution is selected for each assay on each patient
ordered manually. Failure to verify dilution could affect
patient results.
NOTE: If Dil 1 or Dil 2 is selected for HIL, the underline
remains green. However, if the Dil 1 or Dil 2 option is selected
for the HIL reference assay, the selected index assay has a blue
or yellow underline as appropriate.
Option Description
Standard The default dilution configured
Dil 1 Dilution Option 1
Dil 2 Dilution Option 2
Option Line Color
Standard Green
Dil 1 Blue
Dil 2 Yellow
Dil 1 and 2 (Dilutions must be
configured for each assay and will
vary. Cannot see dilution on this
screen— would have to go to assay
file.
!