MODULE 4: RUN OPTIONS
4-24 AEROSET
®
Customer Training Guide
94858-103 — July 2003
5. Select <UNPAUSE> in the Action Area. The RUN OPTIONS
screen is displayed.
NOTE: Do not select <RUN> to reinitiate sampling. If
<RUN> is selected, all samples remaining on the System
with tests still in process are sampled again.
6. Ensure that the Patient option is still selected and that the carrier
number specified when the test was initially ordered is entered in the
Car’r column of the RUN OPTIONS screen.
7. Select <Unpause> in the right-hand column of the RUN
OPTIONS screen to resume sampling.
NOTE: The Operator can select <UnpsCrsl> and the
FastTrack Sampler remains in the PAUSE status.
When the Sample Carousel has Completed Sampling
When the Sample Carousel has completed sampling the Sample
Carousel Movement Indicator Light is not illuminated or blinking. If the
FastTrack™ Sampler is still sampling, the System status is RUN but if
both the Sample Carousel and the FastTrack Sampler have completed
sampling the System status is SAMPLING END, CUVETTE FILL, or
SUSPEND.
• If the System status is CUVETTE FILL or SUSPEND, load the
samples and initiate the run using the steps for a new run.
• If the System status is RUN or SAMPLING END, perform the
following steps:
1. Order the patient samples in the Order Samples
screen.
2. Confirm that the Sample Carousel Movement
Indicator Light is not illuminated or blinking.
3. Select <PAUSE> in the Action Area. The RUN
OPTIONS screen is displayed.
4. Select <Pause> in the right-hand column of the RUN
OPTIONS screen to pause sampling.
5. Wait for the System status to change to PAUSE.
6. Place the samples in the Sample Carousel in the
specified position.