Adding Samples During a Run
AEROSET
®
Customer Training Guide 4-25
94858-103 — July 2003
NOTE: When loading samples during the READY
status, the Sample Carousel begins with
Position A1. When loading samples during the
PAUSE status, the Sample Carousel begins with the
position located at the aspirate position.
7. Select <UNPAUSE> in the Action Area. The RUN
OPTIONS screen is displayed.
CAUTION: Do not select <RUN> to reinitiate sam-
pling. If <RUN> is selected, all samples remaining
on the System with tests still in process are sampled
again.
8. Ensure that the Patient option is still selected and
that the carrier number specified when the test was
initially ordered is entered in the Car’r section of the
RUN OPTIONS Screen.
9. Select <Unpause> in the right-hand column of the
RUN OPTIONS screen to resume sampling.
NOTE: The Operator can select <UnpsCrsl> and
the FastTrack™ Sampler remains in the PAUSE
status.
What’s the Status?
Use the information in the following table to determine which process to
initiate for each status message.
Instrument Status Process
*READY
*CUVETTE FILL
*SUSPEND
1. RUN in Action Area.
2. Check run options.
3. START in Run Options.
*RUN (Sample Carousel)
*SAMPLING END
1. PAUSE in Action Area.
2. Check run options.
3. PAUSE in Run Options.
4. UNPAUSE in Action Area
5. UNPAUSE in Run Options.
RUN (FastTrack) Place the sample carrier to the right of the
Sampler Area on the FastTrack.
!