MODULE 5: RESULT HANDLING
5-6 AEROSET
®
Customer Training Guide
94858-103 — July 2003
•Sample ID
• Patient Name
• Date and Time Ordered
• Patient ID
• Date and Time Completed
• Assay Name (1 through 100)
•Operator
• Comment
• Doctor Name
• Location
The Processing Code column (left-hand column prior to the C/P column)
displays a single-letter code to indicate the status of the completed
samples. The following Processing Codes may be displayed:
The Assay Name columns display the results of the assays ordered for
each sample. If an “*“ is displayed, the result is not complete. If a “C” is
displayed next to the result, an error has occurred for the result. To
review the Result Error Code, access the RESULT screen.
Processing
Code
Status Description
I
Incomplete An error occurred for one or more of the
sample results. This may also indicate that
the sample is an original sample, with rerun
or replicate results.
A
Accepted All results ordered for the sample
completed with no errors.
V
Validated The sample has been validated. This can be
done by the Operator or the System can
auto-validate if configured to do so in the
SYSTEM CONFIGURATION screen.
R
Rerun A rerun was requested for one or more
results on the sample.
L
Linked This sample is a rerun of an original sample
order and is linked to the original sample.
r
Replicate This sample is a replicate of an original
sample order and is linked to the original
sample.