Reviewing Results
AEROSET
®
Customer Training Guide 5-13
94858-103 — July 2003
Numerical Result The result displays unless an error occurred which
prevented the result from calculating. The color of the
result numerical text varies depending on the type of
code associated with it.
• Black - The result is acceptable.
• Brown - The result is outside the reference range
and the appropriate L/H Flag is displayed.
• Red - A Result Error Code has been generated for
the result and the code is displayed.
Result Flag A single-letter flag to indicate when a result is outside
the defined reference range.
• L - Out of range low
• H - Out of range high
Result Error Code If one or more errors occurred for the result, the
highest priority Result Error Code is displayed.
Example:
• SS - Short Sample
• CAL - Calibration Error
• LMP - Lamp Intensity Low
To view all Result Error Codes generated for a result,
select <RxnGraph>.
Measurement Type Indicated by a colored line beneath the assay name.
• Green - Standard
• Blue - Dil 1
• Yellow - Dil 2
• Pink - Reference assay (System ordered assays for
calculated tests and Sample Interference Indices).
Result Information Description
• Show sample with different
measurement types.
• If you order a BUN/Crea ratio,
but only order a BUN, the
instrument will order the
Crea for you and mark it with
a pink underline, but will not
be reported.