The Limited Run Feature
AEROSET
®
Customer Training Guide 5-19
94858-103 — July 2003
5. Select <Order> to the right of the Limited option. The SELECT
ASSAYS TO RUN screen is displayed.
Figure 5.11: SELECT ASSAYS TO RUN Screen
6. Select the assays to be run (the assays to be excluded must not be
selected).
WAR NI NG : Assays with button text that is green in color meet
all required conditions. If the button text is pink or yellow,
further investigation is necessary before performing the assay.
For more information, refer to Confirmation of Assay
Parameter Settings in Section 2, Installation Procedures and
Special Requirements of the AEROSET System Operations
Manual.
7. When assay selection is complete, select <OK> in the upper right-
hand corner of the SELECT ASSAYS TO RUN screen. The RUN
OPTIONS screen displays.
8. Select <Start> in the right-hand column to initiate the run with the
excluded assays.
!