MODULE 5: RESULT HANDLING
5-20 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Deselecting the Limited Run Feature
Perform the following procedure when assays should no longer be
excluded.
1. Select <RUN> from the Action Area. The RUN OPTIONS screen
is displayed.
2. Remove the checkmark (;
) from the Limited option by touching
the box.
3. Select <Start> in the upper, right corner of the RUN OPTIONS
screen to begin sampling. All pending assays are aspirated.
CAUTION: If the limited assay required pretreatment,
load the pretreated sample in the correct carrier position. If
the original sample is not removed and replaced with
the pretreated sample, patient results are affected.
Ensure all samples are present and ready to run when the Limited Run
feature is deselected.
!