Rerunning Results
AEROSET
®
Customer Training Guide 5-25
94858-103 — July 2003
Rerunning Results in the Reserved STAT Position
1. Request the rerun in the DATABASE or RESULT screens as
described in the previous section.
2. Confirm that the correct sample is located in the Reserved STAT
Position.
CAUTION: It is important to confirm the identification of
the sample in the Reserved STAT Position because bar
code labels are not read in this position. Failure to confirm
the identification of the sample could result in sample
misidentification.
3. Select <STAT> in the Action Area. The SELECT ASSAYS FOR
STAT screen is displayed.
4. Enter the Sample ID (must be the same Sample ID displayed in the
DATABASE screen). Do not select any assays.
5. Select <START> in the right-hand column.
!