MODULE 5: RESULT HANDLING
5-24 AEROSET
®
Customer Training Guide
94858-103 — July 2003
5. Select <OK>. All assays on the selected sample(s) are scheduled to
rerun.
NOTE: If the System is not configured for Auto Return,
the Operator must manually move the sample to a
sampling position.
6. To initiate the run, select <RUN> in the Action Area. The RUN
OPTIONS screen is displayed.
7. Select the options desired, then select <Start> in the upper
right-hand corner of the RUN OPTIONS screen.
Rerunning Results from the RESULT Screen
To request that a single result be rerun, the RESULT screen is used.
1. Select <Database> in the Information Access Area, if the screen is
not currently displayed.
2. Select the sample to be reviewed.
3. Select <Result> in the right-hand column. The RESULT screen is
displayed.
4. Select the result that needs to be rerun.
5. Select the measurement type desired.
6. Select <Rerun> in the right-hand column (or press [R] on the
keyboard).
An “R” Processing Code is placed to the left of the result.
7. Select <OK>. In the DATABASE screen the sample also has an
“R” Processing Code and turns pink to indicate that it is pending.
NOTE: If the System is not configured for Auto Return,
the Operator must manually move the sample to a
sampling position.
8. To initiate the run, select <RUN> in the Action Area. The RUN
OPTIONS screen is displayed.
9. Select the options desired, then select <Start> in the upper
right-hand corner of the RUN OPTIONS screen.