EasyManua.ls Logo

Abbott Diagnostics Aeroset - Page 211

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Theory
AEROSET
®
Customer Training Guide 6-7
94858-103 — July 2003
Abs
Results are obtained from the absorbance data based on comparison to
the absorbance of water. End-point assays are represented as absorbance,
and rate assays are expressed as rate of absorbance change/minute.
During calibration, sample blank correction is not performed, even when
sample blank is specified for the end-point assay.
Factor
These assays are typically measuring for enzyme activity. Enzyme assays
are measuring a chromophore (NADH, PNP, etc.) that has a specific
response that varies little between different instruments and reagent lots;
therefore, a single factor is used to calculate all results. Only a reagent
blank is performed for calibration. To calculate results, the reagent blank
absorbance is subtracted from the measured absorbance, and the
difference is multiplied by the factor.
Conc = (measured Abs - Rgt Blank Abs) x Factor
NOTE: The single factor used for the Factor method is entered
in the Calibration page of the Assay Configuration screen. For
further information, refer to Module 8, Configuration.
1-Point Linear
A reagent blank and a single calibrator are measured. A calibration curve
is generated using these two data points.
An example is Sample Interference
Indices, Drugs of Abuse
All enzymes = rates
An example is HDL

Table of Contents