Ordering Controls
AEROSET
®
Customer Training Guide 7-7
94858-103 — July 2003
NOTE: If a QC level is not selected, QC will not be performed
for that run. The selected level applies to all assays ordered for
that QC run.
NOTE: For ICT™ analytes, QC results are generated for all
three ICT assays when any ICT QC is ordered.
Automatic QC
The AEROSET System can be configured to perform automatic QC at
the following times:
After reagent lot change
After reagent cartridge change
NOTE: When configured, automatic QC is performed only
when a reagent lot change or cartridge change occurs during a
run. Automatic QC is not performed when a reagent lot or
cartridge change is detected at the start of a run that is initiated
from the READY status.
NOTE: If automatic QC is expected during a run, ensure the
correct controls are loaded before starting the run.
NOTE: Refer to QC Rule Configuration in Section 2,
Installation Procedures and Special Requirements of the
AEROSET System Operations Manual for information on
Automatic QC options.
NOTE: Controls must be defined on the Calibrator/Control
Carousel for the Automatic QC option to be used.
NOTE: Stability of QC material on the Calibrator/Control
Carousel has not been established.