MODULE 8: CONFIGURATION
8-10 AEROSET
®
Customer Training Guide
94858-103 — July 2003
RUN OPTIONS Screen
The following System Configuration options are defined on the RUN
OPTIONS screen:
Option Description
Priority Patient samples in the Sample Carousel are prioritized
over samples loaded on the FastTrack
™
Sampler
when this option is selected. The default is ON
(selected).
Auto Return The System can be configured to automatically return
samples identified for rerun to the Sample Arm. When
configured, all reruns are presented to the Sample Arm
regardless of whether they are:
• Manually ordered reruns
• Auto Reruns
The default is OFF (unselected).
Car’r Allows the Operator to define the carrier
numbers that are used when ordering samples
to run in the Sample Carousel. Refer to
Ordering Patient Samples in the Sample
Carousel in Section 5, Operating Instructions
of the AEROSET System Operations
Manual.
The default settings are:
•A=1000
•B=2000
•C=3000
•D=4000
•E=5000
•F=6000
Auto Return:
• Will make the sample
carousel turn 1 extra time
• When using this function
with FastTrack, the end of
run carrier (blue) must be
used.
• Everything that precedes
that carrier will be checked
for reruns.
• If position on the FastTrack
has already been used, and
another sample in that
position needs a rerun, the
Auto Return feature will
NOT engage. See Rerunning
Results on page 5-94 of the