System Configuration
AEROSET
®
Customer Training Guide 8-11
94858-103 — July 2003
SYSTEM CONFIGURATION Screen
The following System Configuration options are defined on the
SYSTEM CONFIGURATION screen:
Option Description
Auto Rerun When selected, the System automatically reorders assays
for those samples that meet user specified criteria (Rerun
Rules), on an “assay by assay” basis. To activate the Auto
Rerun feature, this option must be selected and assay-
specific Rerun Rules defined on the Assay Confguration
screen. The default is OFF.
NOTE: Samples are not automatically returned
to the sample arm unless the Auto Return option
is selected on the RUN OPTIONS screen.
NOTE: For information about configuration of
Rerun Rules, refer to the Assay Configuration
section in this Module.
Run Suspend
Time
Allows the Operator to configure how long the System
waits (in minutes) once a run is complete. While in the
SUSPEND status, an abbreviated run initialization is
used (when the next run is initiated). The default setting
is 10 minutes.
Enable Cal Edits Allows the Operator to edit read values
(i.e., absorbances) for calibrators in the ASSAY
STATUS screen. The default setting is ON.
NOTE: If configured to ON, only a Log On
level of “super” can edit calibrator values.
Smart Sampling When selected, the System determines the assay sampling
order to maximize throughput, and minimize the number
of empty cuvettes. The default is ON.
For information about configuration of the
sampling sequence, refer to System
Configuration in Section 2, Installation
Procedures and Special Requirements of the
AEROSET System Operations Manual.
Saving Mode When selected, the System minimizes sample usage. The
over-aspiration volume is aspirated once for the same
sample, when aspirating more than one assay, rather than
each time an assay is aspirated. The default setting is ON.
NOTE: Over-aspiration is an additional volume
of sample aspirated before the actual sample
volume required for analysis. This prevents the
sample from being diluted with System water in
the probe.
Auto Rerun:
To get System to automatically
locate and aspirate reruns
without operator
intervention, auto return must
be ON.