MODULE 8: CONFIGURATION
8-14 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Enable Demog.
Edits in RESULT
When selected, the Operator may edit patient
demographics in the RESULT screen. The Operator
must have a Log On level of “super”. The default
setting is ON.
Characters for L/
H Flags
Allows the Operator to select single characters for the
Low and High Result Flags. The default setting is L and
H.
NOTE: These flags appear next to patient
samples when they are below or above the
configured reference range. These flags also
appear next to QC samples when they are
outside the defined 2 SD range.
Sample Bar*
Codes
Allows configuration that requires the System to match
sample bar codes with the entered sample ID. When
selected, orders are tracked in the database by sample
ID. The default setting is OFF.
NOTE: If the Save by Car’r ID option is
selected, the Sample Bar Code option must be
OFF (deselected).
NOTE: This option can only be edited by an
Abbott Representative.
Fixed Digit* When selected, the System requires the same number of
characters for every sample bar code. The number of
characters expected is defined in the Bar Code Length-
Skip-ID option. The default setting is OFF.
Reagent Bar*
Codes
When selected, the System reads reagent bar code
labels. A scan is initiated by selecting <RGT SCAN> in
the Action Area. The default setting is ON. If a reagent
cartridge does not have a bar code label, the reagent
position must be configured by the operator.
Refer to Reagent Configuration in Section 2,
Installation Procedures and Special
Requirements of the AEROSET System
Operations Manual.
NOTE: This option can only be edited by an
Abbott Representative.
Option Description
Sample Bar* Codes:
• Cannot have Save by Carrier ID
ON if this is ON.
• Must have one or the other
selected.
Fixed Digit*:
• This is not check sum and must
be used in conjunction with bar
code length-Skip-ID.