System Configuration
AEROSET
®
Customer Training Guide 8-15
94858-103 — July 2003
Rgt (reagent) Lo
Alert (#T)
Allows the Operator to configure the System to
alert the Operator when any Reagent Cartridge level
falls below the defined number of tests. The alert
level applies to all Reagent Cartridges. The default
setting is 20 tests. This alert can also be configured
for an individual Reagent Cartridge through the
REAGENTS screen.
Refer to Reagent Configuration in Section 2,
Installation Procedures and Special Requirements
of the AEROSET System Operations Manual.
External
Robotics*
Used with Laboratory Automation Systems. The default
is OFF.
NOTE: This option can only be edited by an
Abbott Representative.
ICT* Allows the System to use the ICT™. The default is ICT.
NOTE: This option can only be edited by an
Abbott Representative.
ScreenSaver
(min)
Allows the Operator to define the number of minutes the
System permits for no screen interaction prior to
initiating the screensaver. The default setting is 10
minutes.
NOTE: External screen savers cannot be loaded
on the AEROSET System.
SmplSeq
(Sampling
Sequence)
Allows override of the System default for assay
sampling order.
NOTE: System level requirements to maximize
throughput cannot be overridden. The following
list indicates these requirements, in the order of
sampling:
Normal assays/ICT™
SmartWash™ pairs
Dilutions
Assays with sample volume > 15 µL
The sampling sequence defined is used in each of the
System level requirements.
Select <SmplSeq> in the right column of the
SYSTEM CONFIGURATION screen.
To add an assay to the sampling sequence, select the
first assay in the unselected (left) column, then select
<Add> in the right column of the Sampling
Sequence dialog window.
Option Description
External Robotics*:
If checked with no external
robotics, it will alarm
continuously.