MODULE 8: CONFIGURATION
8-30 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Reagent Link Allows the R1 and R2
reagent pair to be
linked by position. A
maximum of two sets
of reagent cartridges
per assay are allowed.
The lowest R1 position
is linked to the lowest
R2 position and the
highest R1 position is
linked to highest R2
position. Each time a
reagent scan is
performed, the System
relinks reagents
according to position.
NOTE: If the
Multiple Rgt Ctgs
option is configured
ON, the Reagent Link
option is not
functional— i.e., R1
and R2 cartridges are
not linked.
R1 and R2 cartridges
are not linked. R1
and R2 are utilized
independently.
The linked reagent
pair (R1/R2) to be
used first must be
placed in the lowest
available position.
After the reagent pair
is linked by the
System, the Operator
must verify it is
always utilized as a
pair.
CAUTION:
The R1/R2 pair must
be placed onboard at
the same time.
New reagent cartridges
should be placed in a
higher numbered position
than cartridges already
onboard. R1 and R2 are
utilized independently.
Unit Configuration
Parameter
Description of Parameter Rules for Loading Reagents: If Parameter Is...
ON OFF ON OFF
!