MODULE 12: ACTIVITIES
12-8 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Instrument Set-up
Confirm or edit the following configuration parameters:
1. SYSTEM CONFIGURATION screen:
• Select the Auto Validate option
2. Printer A Configuration screen:
• Result Printing Mode > All
• Select the Auto-Print Calibration Reports option
• Select the Auto-Print Control Reports option
• Select the Auto-Print Patient Reports option
3. Printer B Configuration screen:
• Result Printing Mode > Request
• Select the Auto-Print Control Reports option
• Select the Auto-Print Patient Reports option
4. ONLINE CONFIGURATION screen:
• Protocol Type > Type 1’
• Result Transmission Mode > All
• Deselect the Transmit QC Results option
• Deselect the Use Transmission Check Digit option
• RS-232 Config > 8L-1S-NP and 9600
Begin Your Day
1. Perform the START UP Procedure for 24 hour operation.
2. Perform the Daily Checks.
3. Ensure that all assays are ready to run. If any assay buttons on the
ASSAYS screen are not green, evaluate and resolve the issue
(calibrate, add reagent, etc.).
4. Order and run Start-QC for all assays.