MODULE 12: ACTIVITIES
12-10 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Rerunning and Adding Assays
1. The Emergency Room called and asked to add Sodium, Potassium,
Chloride, and Calcium to Anna Septic.
2. Dr. Dolittle wants the glucose rerun on the STAT sample. To rerun
a sample in the Reserved STAT Position, perform the following
steps:
a. Request the rerun on the DATABASE or RESULT screen as
usual.
b. Ensure the correct sample is still loaded in the Reserved STAT
Position.
c. Select <STAT> from the Information Access Area.
d. Enter the bar code number into the Sample ID field on the
ORDER ASSAYS FOR STAT screen. (Do not order any
assays.)
e. Select <Start> in the right-hand column.
Reviewing Results
When the patient samples are completed, perform the following:
NOTE: All samples without Result Error Codes or rerun results
should be auto validated.
1. Select the rerun result for the STAT sample and validate the sample.
2. Review any incomplete samples and accept or select the results.
3. Validate any samples that have not been auto validated.
4. The Emergency Room explained that their computer is down and
that they need to have printed reports sent. Print Form 2 reports for
the 4 priority samples sent from the ER.