GLOSSARY
AEROSET
®
Customer Training Guide Glossary-5
94858-103 — July 2003
Calibration, Linear Mode
(1-Point Method)
A calibration mode in which a reagent blank and one type of calibrator are
measured, and a calibration curve is generated using these two data points.
Calibration, Linear Mode
(Multi-Point Method)
A calibration mode in which a reagent blank and two to eight calibrators of
different concentrations are measured, and a calibration curve is generated
using these data points.
Calibration, Logit-4 A calibration mode for assays in which the absorbance or absorbance change
increases as the concentration increases.
Calibration, Logit-5 A calibration mode for assays in which the absorbance or absorbance change
increases as the concentration increases. Uses one more parameter than Logit-4.
Calibration, Operator-
Specified Calibration
A calibration that is manually requested.
Calibration, Spline A non-linear calibration mode.