MODULE 2: BASIC RUN
2-16 AEROSET
®
Customer Training Guide
94858-103 — July 2003
6. The reagent cartridge is triangle-shaped to match positions in the
segments. Align the cartridge with the position and load the reagent
cartridge. Seat the cartridge completely.
NOTE: The small reagent cartridge and 20 mL reagent
cartridge require the use of an adapter. Load the adapter
first, ensuring proper alignment. Insert the cartridge in the
adapter and seat the cartridge completely. When loading a
20 mL reagent cartridge, face the bar code label toward the
outside of the Reagent Supply Center. Ensure the reagent
cartridge is completely seated in the adapter.
7. Replace the Reagent Supply Center cover. Verify the cover is seated
by attempting to move it left or right. The cover does not rotate
when seated properly.
8. Select <RGT SCAN> in the Action Area of the Main Display to
update the reagent status.
NOTE: If a new reagent lot or cartridge is scanned, an
Error Log Message displays. The Operator can order a
calibration and load the calibrators. If automatic
calibration is defined for lot or cartridge change, the Error
Log Message indicates the calibrators should be loaded.
NOTE: Reagent information only updates for bar code
labeled Abbott reagents. Non-Abbott reagent information
must be manually entered and verified. Refer to Reagent
Configuration in Section 2, Installation Procedures and
Special Requirements of the AEROSET System
Operations Manual for details.
9. The expiration date of reagents (bar code scanned and manually
configured reagents) is checked when the System is powered ON, a
reagent scan is performed, and when the System date changes.
Reagent expiration dates are monitored by date, so a reagent expires
at midnight on the date of expiration.
NOTE: The expiration date must be configured to
generate an expired (EXP) Result Error Code and Error
Log Message.