EasyManua.ls Logo

Abbott Diagnostics Aeroset - Loading the Patient Samples to Run on the FastTrack Sampler

Default Icon
456 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Basic Steps to Order and Run a Sample
AEROSET
®
Customer Training Guide 2-23
94858-103 — July 2003
Loading the Patient Samples to Run on the FastTrack™ Sampler
WARNING: Potential Biohazard. Refer to Appendix,
Hazards of this manual.
When loading samples, ensure that the following conditions are met:
Prior to loading a tube, verify that the size of the tube is listed as an
acceptable sample container.
Ensure there is sufficient sample volume to run the ordered tests
plus any anticipated reruns.
For primary tubes ensure that there is sufficient
sample volume for the assays to be measured, plus
½ inch (13 mm) of sample above the clot, gel
separator, or plasma/red cell interface. If the specimen
volume is less than ½ inch (13 mm), transfer the
specimen into a sample cup.
For a sample cup used in conjunction with a 100 mm
(height) sample tube, do not fill the specimen in the
cup above the height of the tube.
The following table provides the minimum sample volume required in a
sample cup. The minimum volume required is dependent upon the
largest possible aspiration volume of an assay pair ordered on the
sample.
½ inch
(13 mm)
of sample
Top of clot, gel,
or plasma/red cell
interface
Sample

Table of Contents