Basic Steps to Order and Run a Sample
AEROSET
®
Customer Training Guide 2-25
94858-103 — July 2003
1. Load the samples in the specified Sample Carrier and position. If the
sample tube is bar code labeled, face the label toward the right-hand
side of the Sample Carrier.
NOTE: When the System is configured to read sample bar
code labels, the patient samples may be placed in any
Sample Carrier and position. Refer to Module 8,
Configuration, for information on configuring the System
to read sample bar code labels.
2. Place the Sample Carrier(s) into a Sample Carrier Tray.
3. Place the blue End-of-Run Carrier to the right of the last Sample
Carrier loaded with samples, if the System is configured for auto
return of the FastTrack™ Sampler.
4. Place the Sample Carrier Tray onto the FastTrack Sampler, to the
right of the Sampler Area, so that it is aligned between the gray
positioning marks. The white bump on the track belt should be
centered under the tray.
NOTE: Do not load trays when the red RERUN Indicator
on the Front Display Panel is illuminated.