MODULE 3: DAILY START UP AND CHECKS
3-8 AEROSET
®
Customer Training Guide
94858-103 — July 2003
START UP and SHUTDOWN Procedures should be scheduled based
on instrument usage. Two different usage types are described in the
following sections.
The QCAvg option is not active in the AEROSET System. Refer
to the following table for an explanation of Clear Database
options.
24 Hour Operation
If the AEROSET System is operational for the entire 24 hour period or
idle for less than eight hours per day, the following START UP and
SHUTDOWN options should be performed daily at the intervals
specified below.
START UP (30 min)
These options should be performed once per 24 hour period and require
30 minutes to complete:
• Change Water in Bath
• Wash Cuvettes
• Wash Probes (and Mixers) with 1% Alkaline Wash Solution
• Wash ICT™ with ICT Cleaning Fluid and ICT Reference Solution
• Drain & Fill IRef Cup
• Flush Water Lines
Option Description
Cal, QC Deletes Cal/QC records from the Database and
converts patient records to preserved samples. The text
color on the Database for these samples is green.
Patients Deletes Cal/QC records and any patient records that are
not preserved from the Database.
All Records Deletes all records from the Database.
Cal, QC, QCAvg Deletes Cal/QC records from the Database and
converts patient records to preserved samples. The text
color on the Database for these samples is green.
Patients, QCAvg Deletes Cal/QC records and any patient records that are
not preserved from the Database.
All Rec, QCAvg Deletes all records from the Database.