START UP and SHUTDOWN
AEROSET
®
Customer Training Guide 3-9
94858-103 — July 2003
SHUTDOWN (1 min)
These options should be performed once per 24 hour period and require
one minute to complete:
• Clear Database, All Records
NOTE: The capacity of the Database screen is 8,000
sample records or 30,000 results. The Operator will be
notified by an Error Log Message when the capacity is
depleted to 100 records. When the capacity is totally
depleted the Operator will be notified and no additional
orders will be accepted.
When the System is configured to Save by Carrier
ID (no sample bar codes) Sample Carrier/Positions
cannot be reused and must be cleared using this
procedure.
The Delete function on the Database screen will
not clear the Database.
NOTE: If the Cal, QC option or Cal, QC, QC Avg option
is selected instead of the All Records option to Clear
Database, then:
• C/P, SID, and PID are deleted for complete,
incomplete, and validated results; these
samples become preserved samples.
• A search cannot be performed for a preserved
sample.
• Text color on the DATABASE screen is green
for a preserved sample.
• The C/P, SID, and PID of the preserved samples
can be used for new patient samples.
• <Rxn Graph> cannot be selected to view a
reaction graph for a preserved sample.