DD+DIS150.06E
Repair and Service
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2 CR 30-X Chapter 3.5 / 103
03-2009 Type 5175 / 100/110 Agfa Company Confidential
(15) Open screws at input side of power supply.
(16) Remove power supply.
Figure 249
DANGER:
If the cables at the input side of the power
supply are mounted wrong, the housing of
the digitizer can be connected to high
voltage. Risk of live.
Mount the cables at the power supply input as
indicated on the cables and shown in Figure
250.
Figure 250
ASSEMBLY:
(1) Mount in reverse order.
See IMPORTANT note below.
IMPORTANT:
Connect the cables at the output side as shown
in Figure 251.
bl
ac
k
re
d
Figure 251