How to Measure Power Power Measurement
Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition 101
–<Meas Started>, a trigger is output when the averaging time
period of a measurement begins.
– <Disabled>, the output trigger mode is disabled.
5 Press Enter.
Dual Power Meters - Master and Slave
Channels
For the Agilent 81635A Dual Power Sensor and the Agilent 81619A
Dual Optical Head Interface module, channel 1 is the master channel
and channel 2 is the slave channel. The master and slave channels
share the same software and hardware triggering system. For some
commands, setting parameters for the master channel sets the
parameters for the slave channel. In these cases, you may only set
parameters for the slave channel by setting master channel
parameters.
The parameters listed in Table 8 can only be set or viewed using the
master channel.
Table 8 Parameters that can only be set using the master channel
Parameter
<Zero>
<MinMax mode>
<Reset MinMax>
<Data points>
<Averaging time>
<Range mode>
<Input trigger mode>
<Output trigger mode>