Return Loss Meters Error Messages
Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition 387
Return Loss Meters
error message TCFS Shutter Selftest Failed (error value)
description This test verifies the shutter driver by applying a designated test pattern
error value 25060 TCFS shutter driver electronic fails
action Please return the unit to your local service center for repair
error message E10010: Invalid Caldata
description The Caldata loaded on the module is invalid. Either the caldata ver-
sion number is not compatible with the firmware or the caldata is
corrupt (checksum incorrect)
test value (no parameter)
action Please return the unit to your local service center for repair
error message E10012: Dual ported RAM write/read test failed
description Writing bit patterns and reading them back to/from the dual ported
ram failed
test value (no test value)
action Please return the unit to your local service center for repair
error message E10013: FPGA not correctly initialized
description Loading the FPGA with data failed. This indicates a defective FPGA
or an FPGA access problem
test value (no test value)
action Please return the unit to your local service center for repair
error message E10014: Firmware corrupt, checksum error (a,b)
description Firmware in flash has an invalid checksum. Either the image which
was programmed was corrupt or the flash memory has a failure
test value a is the stored checksum
b is the calculated checksum
action Reload the module firmware. If the failure persists, please return the
unit to your local service center for repair