Performance Tests Performance Test Instructions
344 Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition
33 Using the number keys press the following sequence: 1,0, ., 7, 8, and
9. 10.789 dB should be displayed as the [Cal] value.
34 Press [+/−] several times. The [Cal] field should toggle between
10.789 dB and −10.789 dB.
35 Press Preset, the green hardkey, to return to the Overview Screen.
Module Interaction Test
NOTE The following functional test assumes the use of modules which are
within specification.
1 Setup the test equipment as shown in Figure 199:
Insert the fixed laser source module in slot 1 and the power sensor
in slot 2 . If you use the Agilent 8164A/B Lightwave Measurement
System, insert the back-loadable Tunable Laser module, if available,
in slot 0.
2 Switch on the mainframe and wait until it has booted.
3 Press Preset, the green hardkey, to set all parameters to their
default values.
4 Ensure that [λ], the wavelength of the power sensor module equals
[λ], the wavelength of the laser source in slot 1.
5 Ensure, the attenuation of the laser source is set to 0.00 dB.
6 Move to the power sensor channel, move to [P], press
[Power Unit], and select <dBm> from the menu that appears.
7 Press the key beside the laser output on the module front panel to
switch on the laser source.
8 Note the power reading of the power sensor in the test record. The
Module Interaction Test ends here if you use the Agilent 8163A/B.
NOTE If you use the Agilent 8164A/B, perform steps 9 to 12.
If you use the Agilent 8166A/B, perform steps 9 to 15.
9 Switch your mainframe off, by pressing the Power Key.
10 Move the laser source module from slot 1 to slot 3, and the power
sensor module from slot 2 to slot 4.
11 Connect the output of the laser source in slot 3 to the input of the
power sensor in slot 4.
12 Repeat step items 2 to 8.