Tunable Lase r S ources Error Messages
Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition 383
Selftests
These tests are started after initialization of the instrument. This test
measures defined parameters, and tests them to limits.
error message Initialization Error - module temperature out of range
(please wait a little bit and try again)!
description Module temperature is too high
error value (no error value)
action Doublecheck whether the fan is working and not blocked. Wait some
time and reboot the instrument. If the failure persists, please return the
unit to your local service center for repair.
error message Initialization Error - module temperature too high
(please wait a little bit and try again)!
description Module temperature is too high
error value (no error value)
action Doublecheck whether the fan is working and not blocked. Wait some
time and reboot the instrument. If the failure persists, please return the
unit to your local service center for repair.
error message Main Peltier Temp. Selftest Failed (error value)
description Verifies whether the peltier temperature of the laser diode is within a given
range
error value 13010 Peltier temperature is not in target range
action Please return the unit to your local service center for repair
error message Main Thermo Temp. Selftest Failed (error value)
description Verifies whether the temperature of the thermopositioner is within a given
range
error value 13020 Thermopositioner is not in target temperature range
action Please return the unit to your local service center for repair
error message Main Peltier Curr. Selftest Failed (error value)
description Checks if peltier current for the laser diode peltier is below maximum
error value 13050 Laser peltier current is above maximum value
action Please return the unit to your local service center for repair