Error Messages Tu na ble Lase r S ources
382 Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition
error message Initialization Error (TASK_VERIFY, id1, id2)
description A task could not be started
error value id1 task name
id2 maximum available memory
action Reboot the instrument. If the failure persists, please return the unit to
your local service center for repair
error message Initialization Error (Tcfs Amplifier Calibration failed, error value)
description The calibration of the tuning control feedback system amplifier has
failed
error value 0 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (TcfsAdpEEPROMInit, error value)
description Error reading tuning control feedback system adapter board EEPROM
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (TcfsEEPROMInit, error value)
description Error reading tuning control feedback system EEPROM
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (Tcfs temperature not settled, error value)
description Temperature has not settled at the tuning control feedback system
error value 0 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (TmpCalDataInit, error value)
description Error accessing caldata
error value 1 (not significant)
action Please return the unit to your local service center for repair
error message Initialization Error (UpPiezoEnable, error value)
description Over current
error value 1 (not significant)
action Please return the unit to your local service center for repair