Using Modules Installation and Maintenance
Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition 285
1 Position the module at an unoccupied slot, with the catch at the
bottom front of the module.
2 Insert the module into the slot and onto the tracks. If the module
does not slide in freely, check that you have correctly positioned and
correctly oriented it and that there is no obstruction to its
movement.
3 Apply pressure to the front panel, and push the module as far as it
goes. You hear a small click when the module reaches its installed
position. This is the catch making contact.
How to Remove a Back-Loadable Module
CAUTION Disconnect all electrical and optical connectors before you remove this
module from the instrument, as this can cause damage to the
connectors.
Make sure that the instrument is in stand-by mode, see page 275,
before you remove a module.
1 Untighten the retaining screws, see Figure 189, that secure the
module in the instrument.
Figure 189 Back Panel of Agilent 8164B Lightwave Measurement System
Figure 190 Side View of a Back-Loadable Module
Handle
Retaining
Screws