Error Messages Errors appearing on pop-up menus
378 Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition
Error
One or more of the following reasons may cause this error message:
Error
Sweep Parameter Inconsistent
Channel <slot>.<channel>
firmware revision all
modus stepped sweep
parameter(s) lambda logging on
action switch lambda logging off
firmware revision <= 2.5x
modus continuous sweep
parameter(s) stop wavelength <= start wavelength
sweep time too small (< 0.001s)
sweep time too large (t > 400s)
trigger frequency too high (f > 10kHz)
lambda logging:
- coherence control active
- cycles 1
- trigger step finished
action check settings and correct where appropriate
firmware revision > 2.6
modus continuous sweep
parameter(s) stop wavelength <= start wavelength
sweep time too small (< 0.001s)
sweep time too large (t > 400s)
trigger frequency too high (f > 10kHz)
lambda logging:
- coherence control active
- trigger step finished
action check settings and correct where appropriate

Table of Contents

Other manuals for Agilent Technologies 8166B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8166B and is the answer not in the manual?

Agilent Technologies 8166B Specifications

General IconGeneral
BrandAgilent Technologies
Model8166B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals