A Background to Return Loss Measurement Return Loss Measurement
Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition 175
That is:
where:
The constants t
1
, t
2
, k
1
and k
2
are multipliers giving the proportion of
power transmitted through the coupler from the Input port to the
Output port and from the Output port to the sensor port respectively.
In other words, when optical power is input at the Output port, k
2
times that power is output at the sensor port. It is not necessary to
know the value for these constants, they can be eliminated later.
The constant s is a multiplier giving the scattering factor. The
scattering factor accounts for the directivity of the second coupler,
backscatter in the fiber, and reflections of connectors. The calibration
procedure helps you to eliminate the affect of these on return loss
measurements.
For Return Loss Measurement” on page 154, the reflection factor of
the component is known. Here we refer to the reflection factor as R
Ref
.
This gives the following equation:
For Measuring the Power when there are No Reflections” on
page 172, the value of the reflection factor is zero. This gives the
following equation:
For Measuring the Reflections from the DUT” on page 173, the value
of the reflection factor of the DUT is called R
DUT
. This gives the
following equation:
Pt
2
k
1
k
2
P
src
Rt
2
s M+=
M t
1
P
src
=
c
1
t
2
t
1
----
k
1
k
2
=
c
2
t
2
t
1
----
s =
P
Þ
c
1
M Rc
2
M+=
(1)
P
Ref
c
1
M
Ref
R
Ref
c
2
M
Ref
+=
(2)
P
para
c
2
M
para
=
(3)
P
DUT
c
1
M
DUT
R
DUT
c
2
M
DUT
+=
(4)

Table of Contents

Other manuals for Agilent Technologies 8166B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8166B and is the answer not in the manual?

Agilent Technologies 8166B Specifications

General IconGeneral
BrandAgilent Technologies
Model8166B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals