How to Use a Variable Optical Attenuator module Setting Attenuation and/or Power Levels
Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition 191
Calibration Procedure To calibrate a test setup that utilizes an 81560A/61A or
81570A/71A/73A Attenuator module, first measure the power applied
by your test set up to the DUT input, then store this value as the
Attenuator’s P
ref
parameter.
After calibration, you edit the Attenuator’s P
SET
parameter to set the
power applied to the DUT.
1 Disconnect the DUT input in your test setup and connect it to your
reference powermeter.
Set the λ parameter for the reference powermeter to the source
wavelength or, for a multi-wavelength DWDM signal, the effective
wavelength λ
effective
.
Set an appropriate measurement <Averaging Time>.
2 Display the details screen for the Attenuator module.
Make sure the λ parameter set for the Attenuator module matches
the source wavelength or, for a multi-wavelength DWDM signal,
λ
effective
.
3 If the reference powermeter is hosted by the same mainframe as the
Attenuator module:
Select the P
ref
field,
Press the [PM-> ref ] softkey,
Select the channel used to host your reference powermeter,
Press [OK].
The power value measured by your reference powermeter is copied
to the Attenuator and stored as P
ref
.
4 If the reference powermeter is hosted by another mainframe:
Manually enter the value of P
ref
such that P
SET
equals the power
value measured by your reference powermeter.
5 Disconnect the input to the reference powermeter, then connect it
to the DUT input.
6 Edit the P
SET
parameter to set the power applied to the DUT.
The attenuation, α , displayed is updated to reflect the new P
SET
value.
NOTE Your calibration is valid until there is some change to the power
applied to the input of the Attenuator module. For example, if you
change the optical path to the Attenuator module, you must
recalibrate.

Table of Contents

Other manuals for Agilent Technologies 8166B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8166B and is the answer not in the manual?

Agilent Technologies 8166B Specifications

General IconGeneral
BrandAgilent Technologies
Model8166B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals