Error Messages Optical Switch modules
406 Agilent 8163A/B, 8164A/B, and 8166A/B User’s Guide, Fourth Edition
error message KeySelftest() called without calling KeyInit() before
KeySelftest(): KeyDisable() failed
KeySelftest(): selftestPre() failed
KeySelftest(): key pressed during selftest or hardware problem
KeySelftest(): selftestPost() failed
KeySelftest(): KeyEnable() failed
KeySelftestExecPortF(): wrong interrupt number
KeySelftestPreTPU(): FreeTpultc() failed
KeySelftestExecTPU(): GetTpuDioState() failed
KeySelftestPostTPU(): AllocTpultc() failed
Selftest (stAll, 0) failed
description Selftest problem. Possible causes:
- hardware failure
- key pressed during selftest
- calibration data or firmware incorrect
test value
action Please return the unit to your local service center for repair
error message InitKey(): unsupported keyLocation
KeyAck(): unsupported keyLocation
KeyEnable(): unsupported keyLocation
KeyDisable(): unsupported keyLocation
KeySelftestPre(): unsupported keyLocation
KeySelftestExec(): unsupported keyLocation
KeySelftestPost(): unsupported keyLocation
description Calibtation data incorrect
test value
action Please return the unit to your local service center for repair
error message Preset (0) failed
Preset(): channel X, config Y failed
description Preset failed, probably due to incorrect calibration data
test value X module channel
Y switch configuration number
action Please return the unit to your local service center for repair
error message Error checking firmware CRC!
description The firmware CRC is wrong