Triggering 4
InfiniiVision Oscilloscopes User’s Guide 139
Hex Bus Pattern Trigger
You can specify a bus value on which to trigger. To do this, first define the bus. See “To
display digital channels as a bus” on page 381 for details. You can trigger on a bus value
whether you are displaying the bus or not.
To trigger on a bus value:
1 Press the [Pattern] key on the front panel.
2 Press the Channel softkey and rotate the Entry knob to select Bus1 or Bus2.
3 Press the Digit softkey and rotate the Entry knob to select a digit of the selected bus.
4 Press the Hex softkey and rotate the Entry knob to select a value for the digit.
5 You can use the Set all Digits softkey to set all digits to a particular value.
When a hex bus digit contains one or more don’t care (X) bits and one or more bit with a
value or 0 or 1, the “$” sign will be displayed for the digit.
For information regarding digital bus display when Pattern triggering see “Bus values are
displayed when using Pattern trigger” on page 384.
NOTE
If a digit is made up of less than four bits, then the value of the digit will be limited to the
value that can be created by the selected bits.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals