5 Displaying
178 InfiniiVision Oscilloscopes User’s Guide
Enabling Fast Debug AutoScale
The Fast Debug option changes the behavior of [AutoScale] to let you make quick visual
comparisons to determine whether the signal being probed is a DC voltage, ground, or an
active AC signal. Channel coupling is maintained for easy viewing of oscillating signals.
1 Press [Utility] > Options > Preferences > AutoScale.
2 In the AutoScale Preferences Menu, press Fast Debug to enable or disable the option.
Specifying the Channels Displayed After AutoScale
The Channels softkey selection determines which channels will be displayed on
subsequent AutoScales.
• All Channels — The next time you press [AutoScale], all channels that meet the
requirements of AutoScale will be displayed.
• Only Displayed Channels — The next time you press [AutoScale], only the channels
that are turned on will be examined for signal activity. This is useful if you only want to
view specific active channels after pressing [AutoScale].
Preserving the Acquisition Mode During AutoScale
The acquisition mode is normally switched to Normal when AutoScale is performed.
AutoScale can be set to leave the acquisition mode unchanged if you prefer this option.
Choose Normal to make the oscilloscope switch to Normal acquisition mode whenever
the [AutoScale] key is pressed. This is the default mode.
Choose Preserve to make the oscilloscope remain in the acquisition mode you have
chosen when the [AutoScale] key is pressed.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals