14 Utilities
392 InfiniiVision Oscilloscopes User’s Guide
To restore the oscilloscope to its default configuration
Press the [Save/Recall] key, then press the Default Setup softkey.
The default configuration returns the oscilloscope to its default settings. This places the
oscilloscope in a known operating condition. The major default settings are:
Horizontal normal mode, 100 s/div scale, 0 s delay, center time reference.
Vertical (Analog) Channel 1 on, 5 V/div scale, DC coupling, 0 V position, 1 M
impedance, probe factor to 1.0 if an AutoProbe probe is not connected to the channel.
Trigg er Edge trigger, Auto sweep mode, 0 V level, channel 1 source, DC coupling, rising
edge slope, 60 ns holdoff time.
Display Vectors on, 33% grid intensity, infinite persistence off.
Other Acquire mode normal, [Run/Stop] to Run, cursors and measurements off.
Labels All custom labels that you have created in the Label Library are preserved (not
erased), but all channel labels will be set to their original names.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals