5 Displaying
174 InfiniiVision Oscilloscopes User’s Guide
Tips for Displaying Waveforms
The easiest way to display waveforms is to connect the probes to the signals and press
[AutoScale].
When selecting a particular trigger type, it is helpful to consider whether to use Auto or
Normal trigger mode. See “Auto and Normal Trigger modes” on page 84.
Stabilizing the Display
If you are acquiring a slow signal from a serial bus or another signal that arrives in bursts,
the oscilloscope may Auto-trigger between signal bursts, which will cause the display to
appear unstable. You can solve this problem by selecting the Normal trigger mode. See
“Auto and Normal Trigger modes” on page 84.
If the acquired signal is noisy, there are several steps you can take to stabilize its display.
To reduce the random noise displayed when triggering on a noisy signal, see “Trigger
Noise Rejection” on page 87.