Triggering 4
InfiniiVision Oscilloscopes User’s Guide 159
6 If you select a line # mode, press the Line # softkey, then rotate the Entry knob to
select the line number on which you want to trigger.
7 When using the Generic standard and you select a line # mode or Count:Vertical,
press the Count # softkey and rotate the Entry knob to select the desired count
number.
Listed below are the line (or count) numbers per field for each non-HDTV/EDTV video
standard.
Line numbers for each EDTV/HDTV video standard
TV Triggering Examples
The following are exercises to familiarize you with TV triggering. These exercises use the
NTSC video standard.
Video standard Field 1 Field 2 Alt Field
NTSC 1 to 263 1 to 262 1 to 262
PAL 1 to 313 314 to 625 1 to 312
PAL-M 1 to 263 264 to 525 1 to 262
SECAM 1 to 313 314 to 625 1 to 312
Generic 1 to 1024 1 to 1024 1 to 1024 (vertical)
EDTV 480p/60 1 to 525
HDTV 720p/60 1 to 750
HDTV 1080p/24 1 to 1125
HDTV 1080p/25 1 to 1125
HDTV 1080i/50 1 to 1125
HDTV 1080i/60 1 to 1125
Line Number Represents Count
In Generic mode, the line number represents the number of a count instead of a real line number.
This is reflected in the label in the softkey changing from Line to Count. In the Mode softkey
selections, Line:Field 1, Line:Field 2 and Count:Vertical are used to indicate where the counting
starts. For an interlaced video signal, the counting starts from the rising edge of the first vertical
serration pulse of Field 1 and/or Field 2. For a non-interlaced video signal, the counting starts after
the rising edge of the vertical sync pulse.