Web Interface 9
InfiniiVision Oscilloscopes User’s Guide 277
To connect to the network using Manual mode
1 Get the oscilloscope’s network parameters (hostname, domain, IP address, subnet
mask, gateway IP, DNS IP, etc.) from your network administrator.
2 Connect the oscilloscope to the local area network (LAN) by inserting the LAN cable
into the “LAN” port on the rear panel of the oscilloscope.
3 On the oscilloscope, ensure the controller interface is enabled:
a Press [Utility]&I/O&Control.
b Ensure that LAN is selected. If not, use the Entry knob to point to “LAN” and then
push the Entry knob.
4 Configure the oscilloscope’s LAN interface:
a Press the LAN Settings softkey.
b Press the Config softkey to deselect DHCP and AutoIP.
c Press the Addresses softkey.
d Use the Modify softkey (and the other softkeys and the Entry knob) to enter the IP
Address, Subnet Mask, Gateway IP, and DNS IP values. When you are done, press
the return (up arrow) softkey.
e Press the Domain softkey. Use the Modify softkey (and the other softkeys and the
Entry knob) to enter the Host name and the Domain name. When you are done,
press the return (up arrow) softkey.
f Press the Apply softkey.
To establish a LAN connection (6000L Series)
1 Connect a cable from your site Local Area Network (LAN) to the LAN port on the rear
panel of the oscilloscope.
2 Turn on the oscilloscope power.
3 When the LAN indicator illuminates green, the LAN is connected as configured.
This may take a few minutes.
The oscilloscope will attempt to connect to the LAN using DHCP, AutoIP, then
Manual, in that order (when enabled).
DHCP allows a network DHCP server to assign an address. If a DHCP server is not
present, the oscilloscope moves on to AutoIP.
These connection methods can be enabled or disabled later using the Configure
Network tab in the web interface.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 5000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 5000 Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 5000 Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 5000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals