Serial Decode/Lister 10
InfiniiVision Oscilloscopes User’s Guide 313
Interpreting I
2
S Lister Data
In addition to the standard Time column, the I
2
S Lister contains these columns:
Left Channel — displays the left channel data.
Right Channel — displays the right channel data.
Errors — highlighted in red and marked with an “X”.
NOTE
When the receiver word size is greater than the transmit word size, the decoder fills the
least significant bits with zeros and the decoded value does not match the trigger value.