10 Serial Decode/Lister
324 InfiniiVision Oscilloscopes User’s Guide
2 Specify the SPI signals:
a Press the Settings softkey to access the SPI Settings Menu.
b Press the Signals softkey to access the SPI Signals Menu.
NOTE
If you have already set up an SPI trigger, the signals and their trigger or threshold levels
have already been specified, and you can continue on with step 3.
Changing SPI signals in the Serial Decode setup also changes them in the Trigger setup.