Spectrum Emission Mask (6.6.2.1) 3.1.28.
This chapter describes UL measurement examples where (Modulation, RB) is (QPSK, PartialRB), (QPSK, FullRB),
(16QAM, PartialRB), or (16QAM, FullRB).
[Pass/Fail evaluation limits value setting]
1. Execute SEM_AVG 20 to set the average count of Spectrum Emission Mask to 20 times.
2. Execute TP_SEM5MHZ_1 -13.5 to set the Pass/Fail judgment of Spectrum Emission Mask Frequency Range
0 - 1 MHz.
3. Execute TP_SEM5MHZ_2 -8.5 to set the Pass/Fail judgment of Spectrum Emission Mask Frequency Range 1
- 5 MHz.
4. Execute TP_SEM5MHZ_3 -11.5 to set the Pass/Fail judgment of Spectrum Emission Mask Frequency Range
5 - 6 MHz.
5. Execute TP_SEM5MHZ_4 -23.5 to set the Pass/Fail judgment of Spectrum Emission Mask Frequency Range
6 - 10 MHz.
[(QPSK, PartialRB) measurements]
6. Execute TESTPRM TX_MAXPWR_Q_P to set Test Parameter to TX1 - Max. Power (QPSK/PartialRB).
7. Execute ULRB_POS MIN to set UL RB Position to Min (#0).
8. Execute SWP to measure the Spectrum Emission Mask.
9. Execute SEMPASS? to check that the SEM Pass/Fail judgment is Pass.
10. Execute ULRB_POS MAX to set UL RB Position to Max (#max).
11. Execute steps 8 to 9.
[(QPSK, FullRB) measurements]
12. Execute TESTPRM TX_MAXPWR_Q_F to set Test Parameter to TX1 - Max. Power (QPSK/FullRB).
13. Execute steps 8 to 9.
[(16QAM, PartialRB) measurements]
14. Execute TESTPRM TX_MAXPWR_16_P to set Test Parameter to TX1 - Max. Power (16QAM/PartialRB)).
15. Execute steps 7 to 11.
[(16QAM, FullRB) measurements]
16. Execute TESTPRM TX_MAXPWR_16_F to set Test Parameter to TX1 - Max. Power (16QAM/FullRB).
17. Execute steps 8 to 9.
NOTE 1: The PartialRB allocation UL RB Position is divided as follows:
When Test Frequency is Low range, Max (#max)
When Test Frequency is Mid range, Min (#0) and Max (#max)
When Test Frequency is High range, Min (#0)
NOTE 2: The Pass/Fail evaluation value is initialized as described in TS36.521-1 6.6.2.1.5, 6.6.2.2.5 and
used when the Carrier Frequency is 3 GHz or less. When the Carrier Frequency exceeds 3 GHz, set:
TP_SEM**MHZ_1
TP_SEM**MHZ_2
TP_SEM**MHZ_3
TP_SEM**MHZ_4
as described in TS36.521-1 6.6.2.1.5, 6.6.2.2.5. (** = 1.4, 3, 5, 10, 15, 20).
For the Pass/Fail evaluation values, refer to Chapter 3.7.4 Test Parameter Limit in the operation manual.