Location Registration and Packet Connection 5.2.6.
5.2.6.1. MT8820
1. [SCC] Execute LVL OFF to set SCell output power to off.
2. [PCC] Execute CALLSO to clear call processing.
3. [PCC] Execute CALLSTAT? to confirm the call processing status is 1 (= Idle).
4. Turn on the UE power.
5. [PCC] Execute CALLSTAT? to confirm the call processing status is 6 (= Connected).
Repeat Polling query response when the checked status is not 6 (= Connected).
6. [SCC] Execute LVL ON to set SCell output power to On.
7. [PCC] Execute TPUT_SAMPLE 2000 to set the number of Throughput measurement samples to 2000.
8. [PCC] Execute SWP to perform the Throughput measurement.
9. [PCC] Execute TPUT? PER to read the Throughput measurement result (%).
If an error occurs, the reception state must to be optimized by changing the RMC setting by referring to
Chapter 5.2.5.
5.2.6.2. MT8821
1. Execute CALLSO to clear call processing.
2. Execute CALLSTAT? to confirm the call processing status is 1 (= Idle).
3. Turn on the UE power.
4. Execute CALLSTAT? to confirm the call processing status is 6 (= Connected).
Repeat Polling query response when the checked status is not 6 (= Connected).
5. Execute TPUT_SAMPLE 2000 to set the number of Throughput measurement samples to 2000.
6. Execute SWP to perform the Throughput measurement.
7. Execute TPUT? PER to read the Throughput measurement result (%).
If an error occurs, the DL transmission condition must be optimized by changing the output level or DL RMC
setting of each CC by referring to Chapter 5.2.5.