Maximum Power Reduction (MPR) for HPUE (6.2.3_1) 3.1.4.
The measurement can be performed using the same procedure as Chapter 3.1.3 except the Pass/Fail evaluation
limits value setting.
[Pass/Fail evaluation limits value setting]
1. Execute PWR_AVG 20 to set the average count of power measurement to 20 times.
2. Execute TP_MPR1_LL 26.3 to set TX1 - Max. Power (QPSK/FullRB) Pass/Fail lower limit to 26.3 dBm.
3. Execute TP_MPR1_UL 33.7 to set TX1- Max. Power (QPSK/FullRB) Pass/Fail upper limit to 33.7 dBm.
4. Execute TP_MPR2_LL 26.3 to set TX1 - Max. Power (16QAM/PartialRB) Pass/Fail lower limit to 26.3 dBm.
5. Execute TP_MPR2_UL 33.7 to set TX1 - Max. Power (16QAM/PartialRB) Pass/Fail upper limit to 33.7 dBm.
6. Execute TP_MPR3_LL 25.3 to set TX1 - Max. Power (16QAM/FullRB) Pass/Fail lower limit to 25.3 dBm.
7. Execute TP_MPR3_UL 33.7 to set TX1 - Max. Power (16QAM/FullRB) Pass/Fail upper limit to 33.7 dBm.