Maximum Power Reduction (MPR) for Multi-Cluster PUSCH (6.2.3_2) 3.1.5.
This chapter describes the measurement examples for the following conditions.
First example: Channel Bandwidth = 20 MHz, UL Modulation is 16QAM,
UL Number of RB and Starting RB of Cluster1 is 4,0 respectively, and
UL Number of RB and Starting RB of Cluster2 is 92,8 respectively.
Second example: Channel Bandwidth = 20MHz, UL Modulation is 16QAM,
UL Number of RB and Starting RB of Cluster1 is 92,0 respectively and
UL Number of RB and Starting RB of Cluster2 is 4,96 respectively.
[Pass/Fail evaluation limits value setting for first example]
1. Execute PWR_AVG 20 to set the average count of power measurement to 20 times.
2. Execute TP_MPR2_LL 19.3 to set TX1 - Max. Power (16QAM/PartialRB) Pass/Fail lower limit to 19.3 dBm.
3. Execute TP_MPR2_UL 25.7 to set TX1 - Max. Power (16QAM/PartialRB) Pass/Fail upper limit to 25.7 dBm.
[(16QAM, PartialRB) measurements for first example]
4. Execute TESTPRM TX_MAXPWR_16_P to set Test Parameter to TX1 - Max. Power (16QAM/PartialRB).
5. Execute CHCONFIG PUSCH_MULTI to set Common Parameter - RMC Configuration to PUSCH(Multi
Cluster).
6. Execute ULRB_MULTI 4,0,92,8 to set Common Parameter - UL RMC - 1
st
PUSCH Number of RB to 4, 1
st
PUSCH Starting RB to 0, 2
nd
PUSCH Number of RB to 92, and 2
nd
PUSCH Starting RB to 8.
7. Execute SWP to measure the power.
8. Execute POWER? AVG to read the TX power measurement result.
9. Execute POWERPASS? to check that the TX power measurement Pass/Fail judgment is Pass.
[Pass/Fail evaluation limits value setting for second example]
10. Execute PWR_AVG 20 to set the average count of power measurement to 20 times.
11. Execute TP_MPR2_LL 19.3 to set TX1 - Max. Power (16QAM/PartialRB) Pass/Fail lower limit to 19.3 dBm.
12. Execute TP_MPR2_UL 25.7 to set TX1 - Max. Power (16QAM/PartialRB) Pass/Fail upper limit to 25.7 dBm.
[(16QAM, PartialRB) measurements for second example]
13. Execute TESTPRM TX_MAXPWR_16_P to set Test Parameter to TX1 - Max. Power (16QAM/PartialRB).
14. Execute CHCONFIG PUSCH_MULTI to set Common Parameter - RMC Configuration to PUSCH(Multi
Cluster).
15. Execute ULRB_MULTI 92,0,4,96 to set Common Parameter - UL RMC - 1
st
PUSCH Number of RB to 92, 1
st
PUSCH Starting RB to 0, 2
nd
PUSCH Number of RB to 4, and 2
nd
PUSCH Starting RB to 96.
16. Execute steps 7 to 9.
NOTE 1: The usual Pass/Fail evaluation value is set to the Band 1 default value described in TS36.521-1.
Since the evaluation value differs according to the Band, set the evaluation value described in
TS36.521-1 Table 6.2.3_2.5-1 or Table 6.2.3_2.5-2 at:
TP_MPR2_LL
TP_MPR2_UL
For the Pass/Fail evaluation values, refer to Chapter 3.7.4 Test Parameter Limit in the operation manual.